Technical Information

Cryogen Gas Specifications

CryoTech cryoanalgesic devices are designed to work only with medical grade nitrous oxide (N20) or medical grade carbon dioxide (CO2). A “Certificate of Conformance” is available from your medical gas supplier. The gas cylinders must be fitted with non-siphon valves.

The minimum recommended purity assay for medical grade nitrous oxide is 99.0% with limits to other contents as shown below in a typical “Certificate of Conformance” format:

Test Nitrous Oxide USP Limits
Identification Passes test
Carbon Dioxide 0.03%
Carbon Monoxide 10 ppm
Halogens 1 ppm
Nitric Oxide 1 ppm
Nitrogen Dioxide 1 ppm
Ammonia 0.0025%
Water 150 mg/M3
Air 1.0%
Assay 99.0%